FM/FMR ZSME300000103
6-2 Weight Calibration
Turn on and warm up the machine for 15 to 30 minutes before calibration.
Adjust the calibration switch to “ON”, and the screen will show .
Press or to select .
Procedure
Zero Calibration
a) Ensure nothing on the platter; after being stable, press , and the screen will display
“……. ”. The zero calibration will be complete 5 seconds later.
b) To abort zero calibration, just press .
Weight Calibration
a) Place an object, whose weight is known, on the platter, and input the weight value from
front panel. After being stable, press and the screen will show “……. ”. The
weight calibration will be complete 5 seconds later.
b) To abort weight calibration, just press .
Zero calibration completed
Set calibration
switch to “OFF”
weight
set the